Calibration commands and messages, 5 calibration commands and messages -19, Able – Teledyne 6200A - Sulfides analyzer User Manual

Page 100: 12: rs-232 c, Alibration, Essages, Autocal described in section 6.8.5 and, Te rs-232 operation (see section 6.8.5), Table 6-12, 5 calibration commands and messages

Advertising
background image

TAI Model 6200A SO

2

Analyzer Operator Manual, 02164, Rev. G

Whenever a warning message is reported on the analyzer display, if the RS-232 interface is in
the normal mode (i.e. not in quiet mode) the warning message is also sent to the RS-232
interface. These messages are helpful when trying to track down a problem with the analyzer and
for determining whether or not the DAS reports are actually valid. The warning message format
is for example:

W 194:11:03 0000 SAMPLE FLOW WARNING

The format of a warning command is W command. Examples of warning commands are:

W LIST

List all current warnings

W CLEAR ALL

Clear all current Warnings

Individual warnings may be cleared via the front panel or the command line interface. To clear the
sample flow warning shown above the command would be:

W WSAMPFLOW

6.8.5 CALIBRATION Commands and Messages

Table 6-12: RS-232 Calibration Messages

Message

Description

START ZERO CALIBRATION

Beginning IZS zero calibration

FINISH ZERO CALIBRATION, SO

2

1

=xxxxx PPB

Finished IZS zero calibration

START SPAN CALIBRATION

Beginning IZS span calibration

FINISH SPAN CALIBRATION, SO

2

1

=xxxxx PPB

Finished IZS span calibration

START MULTI-POINT CALIBRATION

Beginning multi-point calibration

FINISH MULTI-POINT CALIBRATION

Finished multi-point calibration

1

Depends on which units are currently selected.


Whenever the analyzer starts or finishes an IZS calibration, it issues a status report to the RS-232
interface. If the RS-232 interface is in the normal mode, these reports will be sent. Otherwise,
they will be discarded. The format of these messages is:

C DDD:HH:MM IIII CALIBRATION STATUS MESSAGE

6-19