Calibration commands and messages, 5 calibration commands and messages -19, Able – Teledyne 6200A - Sulfides analyzer User Manual
Page 100: 12: rs-232 c, Alibration, Essages, Autocal described in section 6.8.5 and, Te rs-232 operation (see section 6.8.5), Table 6-12, 5 calibration commands and messages

TAI Model 6200A SO
2
Analyzer Operator Manual, 02164, Rev. G
Whenever a warning message is reported on the analyzer display, if the RS-232 interface is in
the normal mode (i.e. not in quiet mode) the warning message is also sent to the RS-232
interface. These messages are helpful when trying to track down a problem with the analyzer and
for determining whether or not the DAS reports are actually valid. The warning message format
is for example:
W 194:11:03 0000 SAMPLE FLOW WARNING
The format of a warning command is W command. Examples of warning commands are:
W LIST
List all current warnings
W CLEAR ALL
Clear all current Warnings
Individual warnings may be cleared via the front panel or the command line interface. To clear the
sample flow warning shown above the command would be:
W WSAMPFLOW
6.8.5 CALIBRATION Commands and Messages
Table 6-12: RS-232 Calibration Messages
Message
Description
START ZERO CALIBRATION
Beginning IZS zero calibration
FINISH ZERO CALIBRATION, SO
2
1
=xxxxx PPB
Finished IZS zero calibration
START SPAN CALIBRATION
Beginning IZS span calibration
FINISH SPAN CALIBRATION, SO
2
1
=xxxxx PPB
Finished IZS span calibration
START MULTI-POINT CALIBRATION
Beginning multi-point calibration
FINISH MULTI-POINT CALIBRATION
Finished multi-point calibration
1
Depends on which units are currently selected.
Whenever the analyzer starts or finishes an IZS calibration, it issues a status report to the RS-232
interface. If the RS-232 interface is in the normal mode, these reports will be sent. Otherwise,
they will be discarded. The format of these messages is:
C DDD:HH:MM IIII CALIBRATION STATUS MESSAGE
6-19