Able, 1: z, Alve – Teledyne 6200A - Sulfides analyzer User Manual

Page 83: Peration

Advertising
background image

TAI Model 6200A SO

2

Analyzer Operator Manual, 02164, Rev. G

Table 6-1: Zero/Span Valve Operation

Mode

Description

Reference Section

1.

Front panel operation via CALS and
CALZ buttons.

Calibration Section 7 - Manual Zero/Span
Check.

2.

Automatic operation using AUTOCAL

Setup and use of AUTOCAL is described in
and Section 7.4.

3.

Remote operation using the RS-232
interface

Setup described in Table 6-2 Operation of
AUTOCAL described in Section 6.8.5 and
Section 7 - Calibration.

4.

Remote operation using external contact
closures

Section 7.7 - Automatic operation using
external contact closures. Truth Table 7-8
and Section 9.3.4.3.

Zero/Span valves have 3 operational states:

1. Sample mode. Here both valves are un-energized and sample gas passes through the

sample/cal valve and into the analyzer for analysis.

2. Zero mode. The sample/cal valve is energized to the cal mode. The zero/span valve is un-

energized in the zero mode, thus allowing zero gas to be admitted through the rear panel
bulkhead fitting into the analyzer.

3. Span mode. The sample/cal valve is energized and in the cal mode. The zero/span valve is

energized in the span mode. With both valves on, span gas is admitted through a rear panel
bulkhead fitting into the analyzer.

Zero air and span gas inlets should supply their respective gases in excess of the 700 cc/min
demand of the analyzer. Supply and vent lines should be of sufficient length and diameter to
prevent back diffusion and pressure effects. See Figure 2-3 for fitting location and tubing
recommendations.

Adequate zero air can be supplied by connecting a charcoal scrubber and 5 micron particulate
filter (TAI P/N 000369) to the zero air inlet tubing. The zero air scrubber used in conjunction
with the Zero/Span Valve option provides an inexpensive source of zero air.

6-2

Advertising