Teledyne 6200A - Sulfides analyzer User Manual

Page 86

Advertising
background image

TAI Model 6200A SO

2

Analyzer Operator Manual, 02164, Rev. G

Example 2: to perform zero calibration adjust once per day retarding 15 minutes everyday starting
at 11:30 pm, 12/20/98.

1) MODE: ZERO

2) TIMER ENABLE: ON

3) STARTING DATE: 12/20/98

4) STARTING TIME: 23:30

5) DELTA DAYS: 0

6) DELTA TIME: 23:45

7) DURATION: 15.0 MINUTES

8) CALIBRATE: ON

Example 3: to perform zero-span calibration check once per day at 10:30 pm and zero
calibration adjust once per week starting at 11:30 PM, 12/20/98.

1. Select any one of SEQx and program as example 1.

2. Select any other SEQx and program as following; Always avoid setting two or more

sequences at the same time of the day.

1) MODE:

ZERO

2) TIMER ENABLE: ON

3) STARTING DATE: 12/20/98

4) STARTING TIME: 23:30

5) DELTA DAYS: 7

6) DELTA TIME: 00:00

7) DURATION: 15.0 MINUTES

8) CALIBRATE:

ON

6-5