Page 162: 6200a v, Ariables, Table 9-5, Ensation. see table 9-5, Table 9-5 contains a description of "easy variab

TAI Model 6200A SO
2
Analyzer Operator Manual, 02164, Rev. G
Table 9-5: Model 6200A Variables
No.
Name
Units
Default
Value
Value
Range
Description
0
DAS_HOLD_OFF
MIN
15.0
0.5-20.0
Hold off duration after
calibration or diagnostic
mode.
1 TPC_ENABLE
ON
ON-OFF Temp/Pres
compensation
enable
2 RCELL_SET
o
C
50
30-70
Reaction cell temperature set
point
3 IZS_SET
°C
50
30-70
IZS temperature set point
4
DYN_ZERO
OFF
OFF/ON
Enable to adjust zero
calibration through remote
contact closure
5
DYN_SPAN
OFF
OFF/ON
Enable to adjust span
calibration through remote
contact closure
6
RS232_MODE
Bit Field
8
0-99999
Value is SUM of following
decimal numbers:
1=
enable quiet mode
2=
enable computer mode
3=
enable security feature
4=
enable front panel RS-
232 menus
7
CLOCK_ADJ
Sec.
0
± 60
Real-time clock adjustment
9-19