Able, 1: f, Inal – Teledyne 6200A - Sulfides analyzer User Manual
Page 49: Est and, Alibration, Alues
Advertising

TAI Model 6200A SO
2
Analyzer Operator Manual, 02164, Rev. G
Table 2-1: Final Test and Calibration Values
TEST Values
Observed
Value
Units
Nominal Range
Reference Section
RANGE
PPB
50 - 20,000
STABIL
PPB
0.1 - 2
5.2.1, 9.1.1,
Table 9-1, 9.2.5
PRESS
in-Hg
25 - 35
SAMP FL
CC / MIN
650 ± 10%
PMT
mV
0 - 5000
UV LAMP
mV
2000 - 4000 typical
STR. LGT
PPB
<100
DRK PMT
mV
-50 - +200
DRK LMP
mV
-50 - +200
SLOPE
1.0 ± 0.3
OFFSET mV
<250
HVPS
V
550 - 900 constant
DCPS
mV
2500 ± 200
RCELL TEMP
o
C
50 ± 1
BOX TEMP
o
C 8-50
PMT TEMP
o
C
7 ± 1
IZS TEMP
o
C
50 ± 0.3
Electric Test & Optic Test
Electric Test
PMT Volts
mV
2000 ± 1000
SO
2
Conc
PPB
1000 ± 500
Optic
Test
PMT Volts
mV
2000 ± 1000
SO
2
Conc
PPB
1000 ± 500
(table continued)
2-12