Warning commands and messages, 4 warning commands and messages -18, Able – Teledyne 6200A - Sulfides analyzer User Manual

Page 99: 11: rs-232 w, Arning, Essages, 4 warning commands and messages

Advertising
background image

TAI Model 6200A SO

2

Analyzer Operator Manual, 02164, Rev. G

The T command lists TEST messages. Examples of the T command are:

T LIST

Lists test messages currently enabled with T SET

T LIST ALL

Lists all test messages

T RCELLTEMP

Prints the temperature of the reaction cell

T SO2CONC

Prints SO

2

concentration message

T LAMPRATIO

Prints Lamp Ratio

6.8.4 WARNING Commands And Messages

Table 6-11: RS-232 Warning Messages

Name

Message

Description

WSYSRES

SYSTEM RESET

Analyzer was reset/powered on

WRAMINIT

RAM INITIALIZED

RAM was erased

WSAMPFLOW

SAMPLE FLOW WARNING

Sample flow out of spec.

WSAMPPRESS

SAMPLE PRESSURE WARNING Sample pressure below 15” Hg or above

35” Hg

WUVLAMP

UV LAMP WARNING

UV lamp output is below 600mV or
above 4995mV or PMT above 4995mV

WPMTTEMP

PMT TEMP WARNING

PMT temperature too high or too low

WSHUTTTER

SHUTTER WARNING

Shutter not functioning

WRCELLTEMP

RCELL TEMP WARNING

Reaction cell temp. out of spec.

WBOXTEMP

BOX TEMP WARNING

Box temperature too high or too low

WIZSTEMP

IZS TEMP WARNING

IZS temp. out of spec.

WDYNZERO

CANNOT DYN ZERO

Dynamic zero cal. out of spec.

WDYNSPAN

CANNOT DYN SPAN

Dynamic span cal. out of spec.

WHVPS

HVPS WARNING

HVPS too high or too low

WVFDET

V/F NOT INSTALLED

A/D board not installed or broken

WDCPS

DCPS WARNING

DC power supply output below
2300mV or above 2700mV

6-18