Able, 10: epa z, Alibration – Teledyne 6200A - Sulfides analyzer User Manual
Page 121: Rocedure

TAI Model 6200A SO
2
Analyzer Operator Manual, 02164, Rev. G
Zero calibration procedure.
Table 7-10: EPA Zero Calibration Procedure
Step Number
Action
Comment
1.
Press CAL
The 6200A enters the calibrate mode from sample mode.
Allow the zero gas through the sample port.
2.
Wait 10 min
Wait for reading to stabilize at the zero value.
3.
Press ZERO
To select a zero calibration.
4.
Press ENTR
Pressing ENTR actually changes the equations
5.
Press EXIT
6200A returns to the CAL menu.
1. Adjust the SO
2
concentration to approximately 80% of the URL of the range. The expected
SO
2
span concentration can be determined by measuring the cylinder and diluent flows and
computing the resulting concentration.
Calculate the expected concentration as follows.
[
]
[
]
SO
F
F
SO
gas
total
ga
2
=
×
s
----------------------------------- Equation 7.1
2
where;
[SO
2
]
gas
= cylinder gas concentration, ppm
F
gas
=
SO
2
gas flow rate, cc/min.
F
total
= total(gas + diluent) flow rate, cc/min.
[SO
2
]
= expected SO
2
concentration, ppm
Enter the expected concentration [SO
2
] using the procedure in Table 7-12. The expected span
concentration need not be re-entered each time a calibration is performed unless it is changed.
7-14