3.0 TROUBLESHOOTNG

NO POWER:

Plugged in?
Switched on?
Circuit breaker Tripped?

NOT HEATING:

PV/SV switch to PV. Is it heating?
PV/SV switch to SV. Set point correct?
Socket in place on back of controller?
Check M501TRS wiring diagram Figure 3.1

TRS ANALYZER UNSTABLE:

Leak-check.
(Pressurize and see if pressure falls.
Use soap bubble to find leak.)

EFFICIENCY <90%:

Leaking? Leak-check.
Plugged? Compare flow through and bypassing converter.
Flow too high?
Set-point temperature optimized?
Span gas correct?
Contaminated? Check inside of Teflon tubing.

CONVERTER TEMP UNSTABLE: Perform autotune procedure in Section 2.7.2.