Able, 8: a, Ctivity – Teledyne 6200A - Sulfides analyzer User Manual
Page 118: Atrix for, Alibration, Quipment, Upplies, External contact closures. truth table 7-8

TAI Model 6200A SO
2
Analyzer Operator Manual, 02164, Rev. G
Table 7-8: Activity Matrix for Calibration Equipment & Supplies
Equipment/supplies
Acceptance limits
Frequency and method
of measurement
Action if requirements
are not met
Recorder Compatible
with
output signal of
analyzer; min. chart
width of 150 mm (6
in) is recommended
Check upon receipt
Return equipment to
supplier
Sample line and
manifold
Constructed of PTFE
or glass
Check upon receipt
Return equipment to
supplier
Calibration
equipment
Meets guidelines of
reference 1 and
Section 2.3.2 (Q.A.
Handbook)
See Section 2.3.9 (Q.A.
Handbook)
Return equipment/
supplies to supplier or
take corrective action
Working standard
SO
2
cylinder gas or
SO
2
permeation tube
Traceable to NIST-
SRM meets limits in
traceability protocol
for accuracy and
stability (Section
2.0.7, Q.A.
Handbook)
Analyzed against NIST-
SRM; see protocol in
Section 2.0.7, Q.A.
Handbook
Obtain new working
standard and check for
traceability
Zero air
Clean dry ambient
air, free of
contaminants that
cause detectable
response with the
SO
2
analyzer.
See Section 2.9.2 (Q.A.
Handbook)
Obtain air from another
source or regenerate.
Record form
Develop standard
forms
N/A
Revise forms as
appropriate
Audit equipment
Must not be the
same as used for
calibration
System must be checked
out against known
standards
Locate problem and
correct or return to
supplier
7-11