Able, 8: a, Ctivity – Teledyne 6200A - Sulfides analyzer User Manual

Page 118: Atrix for, Alibration, Quipment, Upplies, External contact closures. truth table 7-8

Advertising
background image

TAI Model 6200A SO

2

Analyzer Operator Manual, 02164, Rev. G

Table 7-8: Activity Matrix for Calibration Equipment & Supplies

Equipment/supplies

Acceptance limits

Frequency and method
of measurement

Action if requirements
are not met

Recorder Compatible

with

output signal of
analyzer; min. chart
width of 150 mm (6
in) is recommended

Check upon receipt

Return equipment to
supplier

Sample line and
manifold

Constructed of PTFE
or glass

Check upon receipt

Return equipment to
supplier

Calibration
equipment

Meets guidelines of
reference 1 and
Section 2.3.2 (Q.A.
Handbook)

See Section 2.3.9 (Q.A.
Handbook)

Return equipment/
supplies to supplier or
take corrective action

Working standard
SO

2

cylinder gas or

SO

2

permeation tube

Traceable to NIST-
SRM meets limits in
traceability protocol
for accuracy and
stability (Section
2.0.7, Q.A.
Handbook)

Analyzed against NIST-
SRM; see protocol in
Section 2.0.7, Q.A.
Handbook

Obtain new working
standard and check for
traceability

Zero air

Clean dry ambient
air, free of
contaminants that
cause detectable
response with the
SO

2

analyzer.

See Section 2.9.2 (Q.A.
Handbook)

Obtain air from another
source or regenerate.

Record form

Develop standard
forms

N/A

Revise forms as
appropriate

Audit equipment

Must not be the
same as used for
calibration

System must be checked
out against known
standards

Locate problem and
correct or return to
supplier

7-11