Calibration using dynamic permeation dilution, Able, 12: epa – Teledyne 6200A - Sulfides analyzer User Manual

Page 123: Alibration, Rocedure

Advertising
background image

TAI Model 6200A SO

2

Analyzer Operator Manual, 02164, Rev. G

Table 7-12: EPA Span Calibration Procedure

Step Number

Action

Comment

1.

Press CAL

The 6200A enters the calibrate mode from sample mode.

2.

Wait 10 min

Wait for readings to stabilize at span values.

3.

Press SPAN

If you change your mind after pressing SPAN, you can still
press EXIT here without spanning the instrument.

4.

Press ENTR

Pressing ENTR actually changes the calculation equations.

5.

Press EXIT

6200A returns to SAMPLE mode.


The analog voltage output should measure approximately 80% of the voltage range selected.
(e.g. 4.00VDC if 0-5V output is selected.) The reading on the front panel display should be equal
to the expected SO

2

concentration entered in the procedure given in Table 7-12 above. See the

Troubleshooting Section 9.2.8 if there are problems. Also see the Calibration Quality Check
procedure Section 7.10.

After the zero and the 80% URL points have been set, generate five approximately evenly
spaced calibration points between zero and 80% URL without further adjustment to the
instrument. Allow the instrument to sample these intermediate concentrations for about 10
minutes each and record the instrument SO

2

response.

Plot the analyzer SO

2

response versus the corresponding calculated concentrations to obtain the

calibration relationships. Determine the straight line of best fit (y = mx + b) determined by the
method of least squares.

After the best-fit line has been drawn for the SO

2

calibration, determine whether the analyzer

response is linear. To be considered linear, no calibration point should differ from the best-fit
line by more than 2% of full scale.

7.8.5.2 Calibration Using Dynamic Permeation Dilution

The procedure for multipoint calibration of an SO

2

analyzer by an SO

2

permeation system are

specified in Code of Federal Regulations.

1

Possible dilution system is shown in Figure 7-3. Permeation chamber temperature must be
maintained within ±0.1

o

C of accuracy. Should allow a minimum of 24 hours of temperature

equilibrium period after power up of the temperature controller. Flow controllers should be
capable of maintaining constant flow rate to within ±1%. Flowmeters capable of measuring flow
rates to within ±2% are required. NIST-traceable soap bubble flowmeter or wet-test meter are
suitable for flow determination. Mixing chamber is made of glass and should provide thorough
mixing of SO

2

gas with dilution air. Output manifold should be sufficient diameter to ensure a

minimum pressure drop and must be vented to create ambient pressure at the manifold.
Permeation tube traceability is established by referencing the permeation device to a NIST-SRM
(refer to Table 7-16).

7-16

Advertising