General Specifications

Digital Display

PC

Display Rate

3 Hz numeric display

3 Hz bar graph display and needle display

Scope

500 divisions

Data Displays

Real time, Scope, Averaging, Statistics, Digital tuning needle, Setup

User input correction
factors

1 multipliers and 1 offsets (7 digits floating point)

Analog Output

amplified signal of the head: 0

– 1.832 volt, ± 4% for wattmeter

± 2% for joulemeter

Energy heads and power heads are offset by approximately 626 mV

Radiometer heads are offset by 1.25V

Positive External
trigger

4.5 to 10 V @ 20 mA, optically isolated.

External trigger
pulse width

0.7 us minimum, no maximum time.

Internet Upgrades

Yes

PC Serial Commands

USB

No missing point
throughput serial
frequency

1000 Hz (Turbo Mode)

See rep rate note about being PC dependant

Dimensions

106 mm(W) x 34 mm (H) x 147 mm (D)

Weight

0.424 kg

Power Supply

USB

1

The 2.5µV bias can induce an error into low power measurements with low sensitivity detectors. It is

essential to use the Zero Offset to rezero the M-LINK before making a measurement in these conditions.
It is always good practice to use the Zero Offset. See section 2.2.

2

Varies with detector head.

3

Including linearity.

4

This is PC dependant. A PC with an Intel Duo Core processor and a 2.4GHz clock speed with

Windows XP can run up to 1300 Hz. A Dell laptop with a Pentium and 1.2GHz clock with Windows XP
can only run at 800Hz. Windows 7 machines may experience lower rates. The application software warns
the user if the application cannot keep up and recommends to reduce the rep rate.