LINK User’s Manual Revision 3.0

2

Unpacking

Each Gentec-EO M-LINK is thoroughly tested and calibrated prior to shipment.

Visually inspect every M-LINK unit after removing it from the shipping containers. If you see any damage,
retain all packaging materials and shipping receipts. Any damage claim should be made promptly to the
transportation company. Notify the nearest Gentec-EO representative concerning the claim, so that any
repair or replacement can be arranged as soon as possible.

You can download the PC M-LINK program from our website. Go to the Downloads section of our
website:

www.gentec-eo.com/downloads

. Click on the file name and download it to your PC. The specific

actions necessary vary between browser and browser settings. After it is transferred, open the file on
your PC and follow the instructions to uncompress and install it.

1.2 Specifications

The following specifications are based on a one-year calibration cycle, an operating temperature of 18 to
28ºC (64 to 82ºF) and a relative humidity not exceeding 80%.

Table 1-1 List of Specifications

Power meter specifications

Power Range

4 pW to 30 kW

Power Scales

[2]

(photo diode head)

1 nW, 3nW, 10nW, 30nW, 100nW, 300nW, 1µW, 3µW, 10µW, 30µW,

100µW, 300µW, 1mW, 3mW, 10mW, 30mW, 100mW, 300mW, 1W, 3W

Power Scales

[2]

(thermal head)

300uW, 1mW, 3mW, 10mW, 30mW, 100mW, 300mW, 1W, 3W, 10W,

30W, 100W, 300W, 1kW, 3kW, 10kW, 30 kW

Pyroelectric in power
mode range (UM-B)

100µW, 300µW, 1mW, 3mW, 10mW, 30mW, 100mW, 300mW

Resolution (digital)

Current scale/3000

Monitor Accuracy

±0.5 % ±2 digits

1

Response Time
(accelerated)

[2]

Head dependant

Statistics

Current value, Max, Min, Average, Std Dev., RMS stability, PTP stability.

Data Storage

Continuous on hard drive