Energy meter specifications

Energy Range

30 fJ to 30 kJ

Energy Scales

[2]

(photo diode head)

300fJ, 1 pJ, 3 pJ, 10 pJ, 30 pJ, 300pJ, 1 nJ, 3 nJ, 10 nJ, 30 nJ, 100nJ,

300nJ, 1 uJ, 3 uJ,10uJ, 30uJ, 100µJ, 300µJ, 1mJ, 3mJ, 10mJ, 30mJ

Energy Scales

[2]

(thermal head)

3mJ, 10mJ, 30mJ, 100mJ, 300mJ, 1J, 3J, 10J, 30J, 100J, 300J, 1kJ, 3kJ,

10kJ, 30kJ

Resolution (digital)

[2]

Current scale/3000

Accuracy

[3]

1.0 %±2 digits < 1 kHz

Default Trigger Level

2 %

Software Trigger Level

0.1% to 99%, 0.1% resolution

Repetition
Frequency

[4]

1.0 kHz for data acquisition in real time with time stamp, no missing point

Frequency
Measurement

Accurate frequency measurement up to 1kHz

0-1000 Hz: 0.1% accuracy

Statistics

Current value, Max, Min, Average, Std Dev., RMS stability, PTP stability,

Pulse #, Repetition Rate, Avg Power.

Data Storage

Continuous on Hard drive