Gentec-EO M-LINK User Manual

Page 34

Advertising
background image

M-

LINK User’s Manual Revision 3.0

28

STR

Starts or stops the data flow.

Example.

Send str1CRLF The M-LINK sends data to the com port. Data is sent at the
pulse rate for Energy, at the chop rate for Radiometers, and at 10Hz for
Power.

Send str0CRLF M-LINK stops sending data and replies Ok.

Data is the hexadecimal data string. See the note below for how to decode
this value.

MAX

Query only. Returns the maximum range index for the head in use.

Example

Send maxCRLF

Reply 27CRLF

MIN

Query only. Returns the minimum range index for the head in use.

Example

Send minCRLF

Reply 20CRLF

Advertising