M-

LINK User’s Manual Revision 3.0

30

POL

Sets the external trigger polarity for energy or radiometric measurement

Example

Send pol1CRLF Trigger polarity is rising edge.

Send pol0CRLF Trigger polarity is falling edge.

PRB

Query only. Returns the head connected flag. If no head is present the
command does not reply. This allows the control software to periodically poll
for a head connection.

Example

Send prbCRLF

Reply NCRLF

N = 1, head is connected. No reply means no head is connected.

AUT

Command only. Enables or Disables Auto Ranging.

Example

Send aut1CRLF Enable Auto Range. Replies Ok.

Send aut0CRLF Disables Auto Range, Replies Ok.

NAM

Query only. Returns the head name.

Example

Send nameCRLF

Reply Head NameCRLF

SSE

Command only. Enables or Disables Single Shot Energy for Thermopile
Heads.

Example

Send sse1CRLF Enable SSE Mode, M-LINK replies Ok.

Send sse0CRLF Disables SSE Mode, M-LINK replies Ok.

If a Thermopile head is not in use, M-LINK replies Thermopile Head Only.