CUR

Query only. Returns the current measurement. The argument is 1 for Energy
Heads, Radiometric Heads, or Radiometer heads in Energy mode, and 0 for
all others.

Example

Send cur1CRLF

Reply DataCRLF

Data is in a hexadecimal data string. See the note below: Data Format used
by M-LINK for how to decode this value.

AVG

Sets or Queries the sampling size of the noise suppression.

Example

Send avg16CRLF sets sampling size to 16 pulses, M-LINK replies Ok.

avgCRLF queries the current average.

Example

avgCRLF

Reply 16CRLF

Valid averages are powers of 2 from 1 to 128. If a non valid argument is sent,
Err is returned and the value is left unchanged.

HVR

Query only. Returns the head version.

Example

Send hvrCRLF

Reply VersionCRLF

SSN

Query only. Returns the head serial number.

Example

Send ssnCRLF

Reply Serial NumberCRLF