Display, The statistics display, Tatistics – Gentec-EO M-LINK User Manual

Page 21: Isplay

Advertising
background image

M-

LINK User’s Manual Revision 3.0

15

2.2.4.15 Display

You can Select the Display mode of your choice. The possible display modes are: Statistics, Graphic,
Histogram, Setup and Needle.

2.2.5 The statistics display

Click on Start to begin the statistics. The Batch Size is the number of measurements the statistics
analyses. Select Continuous if you want to collect statistics continually. If the statistics In Range LED
goes red, one of the points went out of range and the statistics values may be incorrect. In Energy mode
the average power can’t be computed if the rep rate is lower than 1 Hz. The M-LINK can’t measure rep
rates lower than 1 Hz.

Fig. 1-6 The Statistics Display

Advertising