Caution – Gentec-EO S-LINK User Manual

Page 10

Advertising
background image

S-

LINK User’s Manual Revision 2.4

5

3-

USB INTERFACE CONNECTOR :

This interface allows remote control and data transfers between the S-Link and a computer
that has a USB communication port.

4-

EXTERNAL POWER SUPPLY INPUT JACK (S-Link with Ethernet option):

Input voltage required: 5-9 VDC/500 mA. If input voltage is between 15 and 25 volts the
monitor will switch to USB power or turn off. If the input power is above 26 volts either the
internal fuse will blow, or, depending on voltage level and waveform, it may damage the
monitor. When the S-Link is under power the green led will light-up. The S-link will work
under all conditions with 1 HP series high power head and all other supported Gentec-eo
heads. When you connect more than one HP series heads, you need to use the USB
power or a 5 volts- 500mA AC to DC power adapter to power the HP head.

CAUTION

Permanent damage may occur to the optical meter if an external
power supply other than the GENTEC-EO 200960 SPU15A-104 is
used. Please call GENTEC-EO or your local distributor if extra
power supplies are needed for a particular setup.

5-

PROBE INPUT JACK:

The S-Link uses a DB-15 female connector to mate with the detector heads (probes).

The S-Link can take advantage of our Personal wavelength correction

™. It reads the

memory in the Smart Interface connector (version 5 and higher) to provide a wavelength
correction that is based on spectral data measured from that specific detector.

Energy detector heads prior to version 4 have a BNC connector. Contact your vendor for a
hardware upgrade.

Power detectors of version V4 and lower and Energy detectors of version V5 and lower
are not supported
. Photodiodes are not supported.