Gentec-EO S-LINK User Manual

Page 31

Advertising
background image

S-

LINK User’s Manual Revision 2.4

26

*ST1/
*ST2

To display information about current status.

See table 1

*SSX000 To set to Binary or ASCII mode.

Example:

*SS1000: Binary Mode for channel 1

*SS1001: ASCII mode sends All points

*SS1002: ASCII mode sends 1 out of 2 points

*SS1003: ASCII mode sends 1 out of 3 points

*SS1999: ASCII mode sends 1 out of 999 points ( 999 is the maximum )

*SS2001: ASCII mode sends All points for channel 2

Default value is 0 (Binary mode)

Contact Gentec-eo for Binary mode description.

Note: The PC-Slink will not work properly if you set the Slink to ASCII mode.

*SU1

/

*SU2

To make a physical scale up for channel 1 or 2.

You must stop the continuous acquisition (*CS1, *CS2), before sending this
command.

*SD1

/

*SD2

To make a physical scale down for channel 1 or 2.

You must stop the continuous acquisition (*CS1, *CS2), before sending this
command.

*ET1x

/

*ET2x

To Set the External Trigger for Channel 1 or 2.

x= 0 = disable, x = 1= enable

Default value is 0.

Example : To disable the External trigger for Channel 1 enter *ET10 .

Example : To enable the External trigger for Channel 2 enter *ET21 .

You must stop the continuous acquisition (*CS1, *CS2), before sending this
command.

*EM1x /
*EM2x

For Power heads only. To set the energy Mode to Channel 1 or 2.

x= 0 = disable, x = 1= enable

Default value is 0.