LINK User’s Manual Revision 2.4

28

*VNM

Activate the *VNP polling mode, This mode is only available in ASCII mode

*VNP

Deactivate the *CV1 and *CV2 mode (Default mode at power up)

*RG1 /

*RG2

Return the current minimum and maximum range for channel 1 or 2.

The information is coded

The fist 5 byte is the identifier

00000: minimum scale.

00001: maximum scale.

10000: End of transmission.

The last 4 bytes are the scale coded in hex value, use the scale table bellow to
determine the scale

Example

After a RG1 you will receive

“:000000013

:00001001A

:100000000”

What it means

:000000013 : is the minimum scale 13 hex = 19 = 3m

:00001001A : is the maximum scale 1A hex=26=10 (Watt or Joule)

:100000000 : End of transmission.

*SC1xx /
* SC2xx

To set the Scale to Channel 1 or 2.

xx = See Table bellow, you send the value for the scale in decimal, but when making
a status request the S-link send it in Hexadecimal.

You must stop the continuous acquisition (*CS1, *CS2), before sending this
command.

Scale

xx

1p

00

1 picowatt or picojoule

3p

01

3 picowatts or picojoules

10p

02

10 picowatts or picojoules

30p

03

30 picowatts or picojoules

100p

04

100 picowatts or picojoules

300p

05

300 picowatts or picojoules

1n

06

1 nanowatt or nanojoule