LINK User’s Manual Revision 2.4
28
*VNM
Activate the *VNP polling mode, This mode is only available in ASCII mode
*VNP
Deactivate the *CV1 and *CV2 mode (Default mode at power up)
*RG1 /
*RG2
Return the current minimum and maximum range for channel 1 or 2.
The information is coded
The fist 5 byte is the identifier
00000: minimum scale.
00001: maximum scale.
10000: End of transmission.
The last 4 bytes are the scale coded in hex value, use the scale table bellow to
determine the scale
Example
After a RG1 you will receive
“:000000013
:00001001A
:100000000”
What it means
:000000013 : is the minimum scale 13 hex = 19 = 3m
:00001001A : is the maximum scale 1A hex=26=10 (Watt or Joule)
:100000000 : End of transmission.
*SC1xx /
* SC2xx
To set the Scale to Channel 1 or 2.
xx = See Table bellow, you send the value for the scale in decimal, but when making
a status request the S-link send it in Hexadecimal.
You must stop the continuous acquisition (*CS1, *CS2), before sending this
command.
Scale
xx
1p
00
1 picowatt or picojoule
3p
01
3 picowatts or picojoules
10p
02
10 picowatts or picojoules
30p
03
30 picowatts or picojoules
100p
04
100 picowatts or picojoules
300p
05
300 picowatts or picojoules
1n
06
1 nanowatt or nanojoule