Gentec-EO S-LINK User Manual

Page 8

Advertising
background image

S-

LINK User’s Manual Revision 2.4

3

Energy meter specifications S-Link

Energy Range

0.08µJ to 20 kJ

Energy Scales

20 scales: 10uJ, 30uJ, 100µJ, 300µJ, 1mJ, 3mJ,
10mJ, 30mJ, 100mJ, 300mJ, 1J, 3J, 10J, 30J, 100J,
300J, 1kJ, 3kJ, 10kJ, 30kJ

Resolution (digital)

Normal mode: Current scale/4096

Ascii mode: Current scale/32768

Accuracy

3

1.0 %

50 µV < 500 Hz

2.0 %

50 µV 500 Hz to 1.2 kHz

3%

50 µV 1.2 kHz to 6 kHz

6%

50 µV 6kHz to 10 kHz

Average Power accuracy

Frequency Accuracy error + Joulemeter accuracy.

Frequency Accuracy

0.1 to 1000 Hz : 1%±0.1 Hz

5

1000 to 2000 Hz : 2%±0.1 Hz

5

5000 to 10000 Hz : 3%±1 Hz

5

Default Trigger Level

2 %

Software Trigger Level

0.1% to 99.9%, 0.1% resolution

Real Time Data Transfer

10 kHz/Channel in Normal mode no missing point with

USB 2.0 or Ethernet and recommended PC.

Statistics

Current value, Max, Min, Average, Std Dev., RMS
stability, PTP stability, Pulse #, Repetition Rate, Avg
Power

Data Storage

Limited to PC hard drive free space

General Specifications S-Link

Digital Display

Computer screen

Display Rate

3 Hz in Real time mode.

15 Hz in Graphic mode.

Data Displays

Real time, Ratio, Line plot, Histogram, Statistics and

3D Histogram.

User input correction factors

1 multiplier and 1 offset (7 digits floating point)

3

Including linearity.

5

With the recommended PC, no graphic display, usb latency set to 1ms and no other applications

running. The rep rate is much lower over the RS-232 link (5.7 Khz@115200, 480 Hz @9600, the rep rate
is divided by 2 if two channel is used).