1.5.3

The Real time Display

In Power mode

The Current value is updated at 3Hz.

The lower window shows the Scale, current offset value and wavelength. You can change the
scale using the quick arrows beside the scale value. When you change the scale, the acquisition
and all display modes will stop for a few milliseconds.

In Energy mode

The Current value is updated at 3Hz. But all data in the statistics are updated at up to 10 kHz in
real time. Single shot measurement is always good. When you change the scale you might have
a false measurement but the next measurement is accurate. The

re is no need for “Zero offset” in

energy mode because the S-link compensates for an analog drift.

The information bar at the bottom shows the scale and wavelength. The middle box is blank in
energy mode. You can change the scale using the quick arrows beside the scale value. When
you change the scale, the acquisition and all display modes will stop for a few milliseconds.