S-

LINK User’s Manual Revision 2.4

27

Example : To disable the Energy mode for Channel 1 enter *EM10 .

Example : To enable the Energy mode for Channel 2 enter *EM21.

You must stop the continuous acquisition (*CS1, *CS2), before sending this
command.

*NMV

Active the acknowledge mode

This function tells the Slink to send an ACK after each serial commands received

The PC-

slink will not work if you send this command and don’t send the *NMQ, *RST

or don’t cycle the power before trying to make a connection.

*NMQ

Deactivate the *NMV mode (Default mode)

*NA1 /

*NA2

Return the detector name for channel 1 or 2.

*SN1 /

*SN2

Return the detector serial number for channel 1 or 2.

*CV1 /

*CV2

Return the last Slink measurement for channel 1 or 2. This mode is only available in
ASCII mode.

For joulemeter mode it will be the last laser pulse value, for wattmeter mode it will be
the last measurement taken.

Example to get only 1 value

Set the ASCII mode for channel 1 or 2.

*SS1001 or *SS2001

Set the scale for channel 1 or 2.

*SC1XX or *SC2XX, where XX is the scale you want in decimal.

Set the poling mode

* VNM

Start the internal acquisition

*CA1 or *CA2

Request 1 Value for channel 1 or 2.

*CV1 or *CV2