Caution:

United States federal law restricts this device to sale by or on the order of a licensed physician.

Caution:

The AutoPulse is designed to be used only with ZOLL-approved accessories. The AutoPulse
will perform improperly if non-approved accessories are used.

Caution:

Only use ZOLL Batteries specifically designed for use with the AutoPulse. The use of other
batteries may cause permanent damage to the AutoPulse and will void the warranty.

Caution:

Make sure that the LifeBand is

not

twisted before automatic compressions begin.

Caution:

Do

not

submerge the AutoPulse Platform in liquid.

Caution:

Use care while using sharp instruments around the LifeBand.

Caution:

Do not block the vents of the AutoPulse.

Caution:

Do not use the AutoPulse alone as a patient transportation aid.

Caution:

Straps or restraints used for transportation purposes

must not interfere

with the operation of

the AutoPulse. Specifically, straps across the patient’s chest may restrict the compression/
decompression of the chest. In general, strapping schemes must not alter the alignment of the
patient to the AutoPulse.

Warning:

•

If a user advisory or fault cannot be cleared or a system error occurs during active operation,
immediately revert to manual CPR.

•

Do not strap across, or otherwise constrain, the LifeBand. Constraining the movement of the
LifeBand can damage or break the LifeBand.

•

Do not touch the patient while the AutoPulse is analyzing the patient’s size.

•

Failure to properly position a patient, both vertically and laterally with respect to the
AutoPulse Platform, may cause injury to the patient.

•

If you must move or realign the patient, you must press the Stop/Cancel button before
adjustment.

•

Do not place your hands or any other objects on or under the LifeBand while the AutoPulse
is analyzing the patient or during active operation.

•

Operating the AutoPulse on a patient for extended periods of time may result in minor skin
irritation to the patient.

•

Check the patient’s chest rise during ventilation during active operation.