Battery Physical
Table A-4
Battery Specifications
Category
Specifications
Size (L×W×H)
11.5 in. by 3.2 in. by 2.2 in. (29.2 cm by 8.1 cm by 5.7 cm).
Weight
5.1 lbs. (2.3 kg).
Type
Rechargeable Nickel-Metal Hydride (NiMH)
Battery voltage (nominal)
32.4V
Capacity
3200 mAh (typical)
Initial Battery runtime
(nominal patient)
30 minutes (typical)
Maximum Battery charge
time
Less than 4
1
/
4
hours at 77ºF (25ºC)
Battery test-cycle time
Less than 10 hours per test-cycle session; up to three consecutive sessions
possible.
Required replacement interval 100 full charge/discharge cycles.
Note:
The Battery will not operate after 100 full charge/discharge cycles.