ZOLL AutoPulse Rev D Users Guide User Manual

Page 51

Advertising
background image

P/N 11440-001 Rev. 3

Page 3-13

User Guide

Figure 3-13 Mode Change Confirm Display Panel Screen

6.

If 30:2 on-the-fly mode switching has been enabled you may press the gray Menu/Mode switch
button to switch between 30:2 and continuous compressions. If 15:2 on-the-fly mode switching has
been enabled you may press the gray Menu/Mode switch button to switch between 15:2 and
continuous compressions. The current mode is displayed in the upper left corner of the screen. The
words above the gray Menu/Mode switch button indicate the alternate mode that the AutoPulse will
switch to. If there are no words above the gray button then on-the-fly mode switching is not enabled,
and the device will only operate in the current mode and pressing the gray button will have no effect.

Once the gray Menu/Mode switch button has been pressed you will be asked to confirm the mode
switch (see Figure 3-13) by pressing the gray Menu/Mode switch button twice in rapid secession.
A single tone will sound to confirm that the mode change has been accepted.

7.

To access the patient or to pause the AutoPulse for any reason, press the Stop/Cancel button. The
AutoPulse Platform releases the tension on the LifeBand, allowing the user to pull the bands to the
maximum extended position. 10 seconds after the Stop/Cancel button has been pressed a single
audio alert tone will sound. Three audio alert tones will sound 20 seconds after the pause was
initiated. Audio alert tones will sound continuously after 30 seconds into the pause. The tones can
be temporarily disabled (and re-enabled) by pressing the Tone Mute button (refer to section Section
1.4.2.6, “Tone Mute Button,” on page 1-7) if this function is allo
wed in the Administrative Menu.
Pressing the Stop/Cancel button while paused will exit the paused state and the alert tone will be
stopped.

Note:

Opening the bands during active operation will cause the AutoPulse to stop operation
immediately. To restart compressions, re-fasten the Velcro

®

fastener, clear the Fault by

pulling up on the LifeBand and pressing Start/Continue and then follow the normal
operating steps beginning with step 1 on page 3-8.

3 0 : 2

0 0 : 0 4

STOP

COMPRESSING . . .

Confirm

CONTINUOUS