A.7 battery environmental – ZOLL AutoPulse Rev D Users Guide User Manual

Page 76

Advertising
background image

Page A-4

P/N 11440-001 Rev. 3

User Guide

A.7

Battery Environmental

Table A-5

Battery Specifications

Category

Specifications

Operating temperature

+32º to +113ºF (0º to +45ºC) ambient installed in device

Charge temperature

+41º to +95ºF (5º to +35ºC) ambient (68º to 77ºF [20º to 25ºC] preferred)

Storage temperature

-4º to +77ºF (-20º to +25ºC) ambient for less than six months (may require
test-cycle to meet performance characteristics)

+77º to +95ºF (+25º to +35ºC) ambient for less than two months (may
require test-cycle to meet performance characteristics)

Operating altitude

0 to 15,000 ft. (0 to 4,572 m)

Enclosure protection

Meets IP24 per IEC 60529

Shock

Meets IEC 60068-2-27 Basic Environmental Testing Procedures – Shock (50g,
11ms pulse, half sine wave)

Vibration

Meets IEC 60068-2-6 Basic Environmental Testing Procedures (10 to 150 Hz,
10 m/s

2

)

Meets IEC 60068-2-64 Basic Environmental Testing Procedures – Random
Vibration Broad Band – General Requirements (f1:20, f2:2000, ASD 0.05)

Free fall

Meets IEC 60068-2-32 Basic Environmental Testing Procedures – Free Fall –
Procedure 1.

Table A-6

Battery EMI/EMC Specifications

Title

Standard

Electrostatic discharge

IEC 61000-4-2, Level 3

Radiated emissions

CISPR 11/EN55011, Group 1, Class A
FCC part 15, Class A

Advertising