Changes to the Communication Specification
None.
Default Parameters
‘Prompt for Balance’ option will default to enabled.
Known Issues
Rear Service Panel blank screen
In some instances on a FT or RT terminal with a rear service panel, a blank screen on the rear
service panel or the touch screen being unresponsive has occurred after version 1.8.5 software
has been loaded. To resolve the issue, the technician should perform a restart or reset the
terminal from the front panel in management functions.
Revision History
Date Version
Description
of
Change
May 29, 2007
1.1
Added known issue about the rear service panel
April 10, 2007
1.0
Initial Release