The following parameters have been reset.
Function Parameter
GENERAL
-
OPERATOR = Default
-
LANGUAGE = ITALIAN
-
CONTACT VOLTAGE = 50 V
-
NOMINAL VOLTAGE = 230 V
-
SAFETY INPUT = ENABLED
-
SOUND = ON
R
PE-2WIRE
-
Im
NOM
= 200 mA
-
LIMIT value (200 mA) = 0.30
-
MODE = MANUAL
-
CAL (200 mA) = 0.00
-
LIMIT value 1 (200 mA) = 0.30
-
LIMIT value 2 (200 mA) = 1.00
-
LIMIT value 3 (200 mA) = 5.00
-
LIMIT value 4 (200 mA) = 50.0
-
LMIT mode (25 A) = STANDARD
-
LIMIT value (25 A, STANDARD limit mode) = 0.30
-
LIMIT value 1 (25 A, STANDARD limit mode) = 0.30
-
LIMIT value 2 (25 A, STANDARD limit mode) = 1.00
-
LIMIT value 3 (25 A, STANDARD limit mode) = 5.00
-
LIMIT value 4 (25 A, STANDARD limit mode) = 10.0
-
LENGTH = 2 m
-
LENGTH 1 = 2 m
-
LENGTH 2 = 3 m
-
LENGTH 3 = 10 m
-
LENGTH 4 = 100 m
-
SECTION = 1 mm
2
-
SECTION 1 = 1 mm
2
-
SECTION 2 = 2.5 mm
2
-
SECTION 3 = 10 mm
2
-
SECTION 4 = 35 mm
2
-
MAT. = Cu
-
Z
LINE
= 0.100
-
Z
LINE
1 = 0.100
-
Z
LINE
2 = 0.300
-
Z
LINE
3 = 0.500
-
Z
LINE
4 = 1.000
-
PROTECTION = MCB B
-
I
N
(any protection) = 6 A
-
I
N
(any protection) 1 = 6 A
-
I
N
(any protection) 2 = 16 A
-
I
N
(any protection) 3 = 25 A
-
I
N
(any protection) 4 = 32 A
-
TIMER = 3 s
-
TIMER 1 = 3 s
-
TIMER 2 = 10 s
-
TIMER 3 = 30 min
-
TIMER 4 = 60 min
-
CAL (25 A) = 0.000