Fuse replacement (fuse F1, F2)

The two fuses are general ones for the tester and protect internal circuitry in POWER, R

PE

and

DIELECTRIC measurements.

In case mains switch pilot lamp (5) does not illuminate after connecting the Machinery-Tester
FULLTEST 3 to mains outlet and switching on the mains switch and neither the LC-Display (24) shows

any indication, it is very likely mains fuse F1 (4) or F2 (3) or both to be blown.

To replace the fuse proceed as follows:

1) Open the fuse holder F1 (4) and F2 (3) by using an appropriate screwdriver.
2) Remove the defective fuse and replace it with a new one (T16A/250V, 5×20 mm).

3) Replace the fuse holder.

Fuse replacement (fuse F3 for LOOP, RA and RCD functions)

F3 (FF 12.5 A / 500 V, 6.3 × 32 mm) has blown if:

• Text FUSE F3 appears on the display in LOOP, R

A

or RCD function.

To replace the fuse proceed as follows:

1) Unlock corresponding fuse holder (2) by using an appropriate screwdriver

2) Remove the defective fuse and replace it with a new one (FF 12.5 A / 500 V, 6.3 × 32 mm).
3) Lock the fuse holder again.

Fuse replacement (fuse F4 for RPE function)

Fuse F4 (F20A/500V 6.3 x 32 mm) has blown if message FUSE F4 appears on the display in R

PE

function.


To replace the fuse proceed as follows:


1) Unlock corresponding fuse holder F4 (15) by using an appropriate screwdriver

2) Remove the defective fuse and replace it with a new one.
3) Replace the fuse holder.

☞

If any fuse blows several times (for example in case of operating

error) the instrument must be sent in to the service department in

order to be checked.