HT instruments FULLTEST3 User Manual
Page 88

FULLTEST
3
EN - 87
PHASE SEQUENCE (PHASE
SEQ
)
Display range
U
L1/2
, U
L2/3
, U
L3/1
(V)
Resolution
(V)
Accuracy
Overvoltage protection
360
460
1
± (2% rdg. + 2 dgt)
CAT III 300 V
Display of test result:
1.2.3 (right) or 2.1.3 (left) or 1.1.X (not defined)
TRMS Clamp Current (I
CLAMP
)
Range
Display range
Resolution
Accuracy
(w/o clamp error)
Overvoltage protection
1000 mA
0.0
99.9 mA
0.1 mA
± (3% rdg. + 3 dgt)
One measurement
terminal grounded
100
1000 mA
1 mA
100.0 A
0.00
9.99 A
0.01 A
10.0
100.0 A
0.1 A
1000 A
0.0
99.9 A
0.1 A
100
1000 A
1 A
Input voltage range:
0
1 V AC
Input resistance:
1 M
Frequency range:
40 Hz
100 kHz (characteristics according to
DIN EN 61010-1 Annex A, fig. A.1 and VDE 0411 Part 1)
Clamp type:
AC, output voltage 1 V / measuring range, type
HT96U (measuring ranges 1 A, 100 A, 1000 A)
LIM value (1000 mA range):
Adjustable 0.1 ... 99.9 mA, 100 ... 1000 mA
LIM value (100.0 A range):
Adjustable 0.1 ... 100.0 A
LIM value (1000 A range):
Adjustable 1 ... 1000 A