Part A System installation and start-up Sect. 1.3.4

05/2002

IFC 210 E

11


Recommended length of signal cable A (Type DS) and B (Type BTS)

Primary head

Meter size

Signal cable

DN mm

inch

A B

AQUAFLUX F

10

- 1600

3

/

8

-

64

A1

B1

ECOFLUX IFS 1000 F

10
25

-
-

15

150

3

/

8

1

-
-

½

6

A4
A3

B3
B2

ALTOFLUX IFS 2000 F

150

-

250 6

-

10

A1

B1

ALTOFLUX IFS 4000 F

10
200

-
-

150

1600

3

/

8

8

-
-

6

64

A2
A1

B2
B1

PROFIFLUX IFS 5000 F

2.5
25

-
-

15

100

1

/

10

1

-
-

½

4

A4
A2

B3
B2

VARIFLUX IFS 6000 F

2.5

-

15

1

/

10

- ½

A4

B3

25

-

80 1

-

3

A2

B2

ALTOFLUX M 900

10

-

300

3

/

8

-

12

A1

B1


Diagram A

Diagram B


Field current cable C

Length L

Cross section A

F

(Cu), minimum

0 – 150 m

5 – 500 ft

2 x 0.75 mm

2

/ 2 × 18 AWG

150 – 300 m

500 – 1000 ft

2 x 1.50 mm

2

/ 2 × 14 AWG

300 – 600 m

1000 – 1900 ft

4 x 1.50 mm

2

/ 2 × 12 AWG