KROHNE IFC 210 E-EEx EN User Manual

Page 9

Advertising
background image

Part A System installation and start-up Sect. 1.3.2

05/2002

IFC 210 E

9

1.3.2 Stripping (preparation) of signal cables


Please note: The numbers in the drawings designate the stranded drain wires of signalling cables
A and B, see sectional drawings in Sect. 1.3.1.

primary head

Converter

Length primary

head

Length Converter

mm

inch

mm inch

a 90 3.60

a 40 2.80

b 8

0.30

b

10 0.30

c 25 1.00

d

8

0.30

d

5

0.30

e

70 2.80

e 20 2.00


Signal cable A (type DS),

double shielding

for primary head

Signal cable A (type DS),

double shielding

for IFC 210 E Converter



Signal cable B (type BTS),

with triple shielding (bootstrap)

for primary head

Signal cable B (type BTS),

with triple shielding (bootstrap)

for IFC 210 E Converter



Customer-supplied materials
V

Tin-coat all stranded drain wire ends!

W

Insulation tubing (PVC), Ø 2.0-2.5 mm (Ø 1”)

X

Heat-shrinkable tubing or cable sleeve

Y

Wire end sleeve to DIN 41 228: E 1.5-8

Z

Wire end sleeve to DIN 41 228: E 0.5-8


Signal cable B

bending radius

50 mm (

2”)

bending radius

50 mm (

2”)

Signal cable A

bending radius

50 mm (

2”)

bending radius

50 mm (

2”)

Advertising
This manual is related to the following products: