Sect. 6.4 Part C Special applications, functional checks, service, and order numbers

IFC 210 E

05/2002

62

6.4

Pulsating flow

Important, please note!

•

Observe the directions and regulations and the electrical data specified in the

EC type test certificate.

•

In addition to the regulations for heavy-current installations (VDE 0100),
pay particular attention to the requirements specified in EN 60079-14 “Electrical
equipment in hazardous areas”
.

•

Assembly, installation, commissioning and maintenance work may only be carried
out by ”personnel trained in explosion protection” !



Application
downstream of positive-displacement pumps (reciprocating or diaphragm pumps) without
pulsation dampener

Operator control of the signal converter for the new settings, see Sect. 4 and 5.

To change settings

•

Fct. 3.2 Field Frequency (change magnetic field frequency)

−

Stroke frequency less than 80 strokes/min (at max. pump lift): do not change setting.

−

Stroke frequency 80 – 200 strokes/min (at max. pump lift): change setting to 1/2,

only practical with IFS 5000 F (DN 2.5-100 /

1

/

10

”-4”)

and IFS 4000 F (DN 10, 15, 50-100 /

1

/

10

”,

1

/

2

”, 2”-4”),

Please consult factory where other types and meter sizes are concerned.

−

Please note: given stroke frequencies close to the tripping point of 80 strokes/min,

additional measuring errors of approx. ± 0.5% of the measured value may occasionally
occur.

•

Fct. 3.06 Application (adjust overload point of the A/D converter to the application) Change
setting to “pulsating”.

•

Fct. 1.04 Display (change display presentation of flow)
Change setting to “Trend” to allow better assessment of display unsteadiness.

•

Fct. 1.02 Timeconstant (change time constant)

−

Set to “All” and time (t) in seconds.

1000

−

Recommended:

t [Sec] =

min. strokes / min

−

Example:

min. number of strokes in operation = 50 strokes/minute

1000

t [Sec] =

50 / min

= 20 Sec

With this setting, the residual ripple of the display will amount to
approx.

±

2% of the measured value. Doubling the time constant will

reduce the residual ripple by a factor of 2.