Sect. 5.5

Part B IFC 210 Signal converter

IFC 210 E

05/2002

40

5.5

Display (Fct. 1.04)

Fct. 1.04 Display
Press

→

key.

Contrast = set required contrast

•

Range adjustable from +15 (high contrast) to –15 (= low contrast)

Change setting with

↑

or

↓

key,

Press

↵

key to transfer to subfunction “flow“.

Flow = set required flow unit and number format

•

Auto

(= exponent notation)

•

# . ####

•

## . ##

•

### . #

Change setting with

↑

or

↓

key,

•

####

Transfer to Unit Selection with

→

key.

•

m

3

/hr

(cubic metres per hour)

•

Liter/Sec

(litres per second)

•

US Gal/min

(US gallons per minute)

•

User-defined unit, factory setting = Liter/hr (litres per hour) or US MGal/day , see Sect. 5.17

Select with

↑

or

↓

key.

Press

↵

key to change to subfunction “totalizer”.



Totalizer = set required totalizer unit and number format

•

Auto

(= exponent notation)

•

# . ######

•

## . #####

•

### . ####

•

#### . ###

•

##### . ##

•

###### . #

Change setting with

↑

or

↓

key,

•

#######

Transfer to Unit Selection with

→

key.

•

m

3

(cubic metres)

•

Liter

(litres)

•

US. Gal

(US gallons)

•

User-defined unit, factory setting = Liter (litres) or US MGal/day , see Sect. 5.17

Select with

↑

or

↓

key.

Press

↵

key to change to subfunction “messages”.