)XQFWLRQDOFKHFNV

7.1

Zero check

•

Set zero flow in the pipeline, but make sure that the measuring tube is completely filled
with fluid.

•

Switch on the system and wait 15 minutes.

•

Press the following keys for zero measurement:

Key Display

Description

→

If “Yes” set under Fct. 3.04 ENTR Y CODE, key in

9-keystroke CODE 1 now:

→

→

→

↵

↵

↵

↑

↑

↑

Fct. 1.00

Operation

2x

↑

Fct. 3.00

Installation

→

Fct. 3.01

Language

2x

↑

Fct. 3.03

Zero Point

→

back

↑

enquire

↵

0.00

- - - - - / - - -

Zero measurement in progress, duration approx. 50 seconds.

When flow > 0 WARNING notice appears, confirm with

↵

key.

Store No

If new value not to be stored, press

↵

key (3x) 4x = return to

measuring

mode.

↑

Store Yes

↵

Fct. 3.03

Zero Point

Store new zero value.

(2x) 3x

↵

- - - - - - -

- - - - - / - - -

Measuring mode with new zero.


7.2

Test of measuring range Q

•

For this test a measured value can be simulated in the range of -110 to +110 percent of Q

100%

(full-scale range set, see Fct. 1.01 Full Scale).

•

Switch on the system..

•

Press the following keys for this test::

Key Display

Description

→

If “Yes” set under Fct. 3.04 ENTR Y CODE, key in

9-keystroke CODE 1 now:

→

→

→

↵

↵

↵

↑

↑

↑

Fct. 1.00

Operation

↑

Fct. 2.00

Test

→

Fct. 2.01

Test Q

→

sure

no

↑

sure

yes

↵

0

%

Current, pulse and status indication outputs indicate the

corresponding values.

± 10

%

↑

± 50

%

Select using

↑

or

↓

key

± 100

%

± 110

%

↵

Fct. 2.01

Test Q

End of test, actual measured values again present at outputs.

(2x) 3x

↵

- - - - - - -

- - - - - / - - -

Measuring mode