5.13 Language (Fct. 3.01)

Fct. 3.01 Language
Press

→

key.

Select language for texts in display

•

D

(German)

•

GB/USA

(English)

•

F

(French)

•

others on request

Select with

↑

or

↓

key.

Press

↵

key to return to Fct. 3.01 Language.


5.14 Zero check (Fct. 3.03)

•

Set zero flow in the pipeline, but make sure that the measuring tube is completely filled
with fluid.

•

Switch on the system and wait 15 minutes.

•

Press the following keys for zero measurement:

Key Display

Description

→

If “Yes” set under Fct. 3.04 ENTR Y CODE, key in

9-keystroke CODE 1 now:

→

→

→

↵

↵

↵

↑

↑

↑

Fct. 1.00

Operation

2x

↑

Fct. 3.00

Installation

→

Fct. 3.01

Language

2x

↑

Fct. 3.03

Zero Point

→

back

↑

enquire

↵

0.00

- - - - - / - - -

Zero measurement in progress, duration approx. 50 seconds.

When flow > 0 WARNING notice appears, confirm with

↵

key.

Store No

If new value not to be stored, press

↵

key (3x) 4x = return to

measuring

mode.

↑

Store Yes

↵

Fct. 3.03

Zero Point

Store new zero value.

(2x) 3x

↵

- - - - - - -

- - - - - / - - -

Measuring mode with new zero.


5.15 Determine Entry Code (Fct. 3.04)

Fct. 3.04 Entry Code
Press

→

key.

Choice

•

No

(no code, enter setting mode with

→

key)

•

Yes

(enter setting mode with

→

key and Code 1:

→

→

→

↵

↵

↵

↑

↑

↑

)

Select with

↑

or

↓

key.

Press

↵

key to return to Fct. 3.04 Entry Code.