6 reset totalizer and cancel error messages, 7 examples of setting the signal converter – KROHNE IFC 210 E-EEx EN User Manual

Page 37

Advertising
background image

Part B IFC 210 Signal converter Sect. 4.6

05/2002

IFC 210 E

37

4.6

Reset totalizer and cancel error messages


Cancel error messages in RESET / QUIT menu

Key Display

Description

- - - - - - -

- - - - - / - - -

Measuring mode

↵

Code 2

- -

Key in entry code 2 for Reset / Quit
menu:

↑

→

↑

→

Error Quit ....

Menu for error acknowledgement

...

No

Do

not delete error messages,

press

↵

twice = return to measuring mode.

↑

...

Yes

Delete error messages

↵

Reset totalizer …

↵

- - - - - - -

- - - - - / - - -

Return to measuring mode


Reset totalizer(s) in RESET / QUIT menu

Key

Display

Description

- - - - - - -

- - - - - / - - -

Measuring mode

↵

Code 2

- -

Key in entry code 2 for Reset / Quit
menu:

↑

→

↑

→

Error Quit ....

Menu for error acknowledgement

↵

Reset totalizer …

Menu for resetting totalizer

→

...

No

Do

not delete error messages,

press

↵

twice = return to measuring mode.

↑

... Yes

Reset totalizer

↵

- - - - - - -

- - - - - / - - -

Return to measuring mode


4.7

Examples of setting the signal converter


As an example the cursor, flashing part of display, is shown below in bold type.

•

Change measuring range of current output and value for error messages (Fct. 1.05):

•

Change measuring range from 04-20 mA to 00-20 mA

•

Change value for error messages from 0 mA to 22 mA

Key Display

Description

→

If “Yes” set under Fct. 3.04 Entry Code, key in the

9-keystroke Code 1 now:

→

→

→

↑

↑

↑

↵

↵

↵

Fct. 1.00

Operation

→

Fct. 1.01 Full

Scale

4

×

↑

Fct. 1.05

Current output I

→

Function

→

↵

Range

I

→

04-20

mA

Old current range

2

×

↑

00-20

mA

New current range

2

×

↵

I

Error

→

0

mA

Old value for error messages

↑

22 mA

New value for error messages

↵

Fct. 1.05

Current output I

↵

Fct. 1.00

Operation

↵

Store Yes

↵

- - - - - - -

- - - - - / - - -

Measuring range with new data for the current output


Advertising
This manual is related to the following products: