KROHNE IFC 210 E-EEx EN User Manual

Page 34

Advertising
background image

Sect. 4.4

Part B IFC 210 Signal converter

IFC 210 E

05/2002

34

Fct.

Display-Texts

Description and settings

Value P

Pulse value has been changed.

With the old pulse values the output frequency (F)

would have been exceeded or not reached.

P

min

= F

min

/ Q

100%

P

max

= F

max

/ Q

100%

Check new values!

Gk Value

Set primary constant GK

see primary head nameplate.

Range:

• 1.0000 - 9.9999

Press

↵

key to transfer to subfunction “Field Frequency”.

Field Frequency

Set Magnetic field frequency (f

Net

= power frequency)

•

1

/

2

×

f

Net

•

1

/

6

×

f

Net

•

1

/

18

×

f

Net

•

1

/

36

×

f

Net

Select with

↑

key.

Press

↵

key to transfer to subfunction “Flow Direction”.

for DC devices to subfunction „Line Frequency“

Line Frequency

Normal line frequency in your country

This function is only provided for units with DC power supply

• 50 Hz • 60 Hz

Select with

↑

key.

Press

↵

key to transfer to subfunction “Flow Direction”.

Flow Direction

Define flow direction (in F/R mode: forward flow).

• + Direction

• - Direction

Select with

↑

key.

Press

↵

key to return to Fct. 3.02 “Flowmeter”.

3.03

Zero Point

Zero calibration

• return

(quit function without making change)

• calculate

(calculate new zero value)

Press

↵

key to start, duration approx. 15-90 seconds.

Save new value, select with

↑

key:

• save no

(do not save zero value)

• save yes

(save new zero value)

• change

(change zero value manually)

Press

→

key to transfer to unit selection:

•

m

3

/h • Liter/s • US Gal/min

• any unit, see Fct. 3.05 (factory setting: Liter/h)

Press

→

key to transfer to number setting.

Value may be max. 10% of Q

100%

Press

↵

key to return to Fct. 3.03 “Zero Point”.

3.04

Entrycode

Entry code required to enter setting mode?

• NO (= entry with

→

only)

• YES(= entry with

→

and Code 1:

→

→

→

↵

↵

↵

↑

↑

↑

)

Press

↵

key to return to Fct. 3.04 “Entrycode”.

3.05

User Units

Set any required unit for flowrate and counting

Text Volume

Set text for required flowrate unit (max. 5 characters)

Factory-set: Liter or MGal .

Characters assignable to each place: • A-Z, a-z, 0-9, or — (= blank character)

Press

↵

key to transfer to subfunction “Factor Volumen”.

Factor Volume

Set conversion factor (FM) for volume

Factory set 1.00000 for Liter or 2.64172E-4 for US MGal

(exponent notation, here: 1x 10

3

or 2.64172x10

-4

).

Factor FM = volume per 1m

3

.

Setting range • 1.00000 E-9 to 9.99999 E+9 (= 10

-9

to 10

+9

)

Press

↵

key to transfer to subfunction “Text Time”.

Text Time

Set text for required time unit (max. 5 characters)

Factory-set: hr.

Characters assignable to each place: • A-Z, a-z, 0-9, or — (= blank character)

Press

↵

key to transfer to subfunction “Factor Time”.

Factor Time

Set conversion factor (F

T

) for time

Factory-set: 3.60000 E+3 for hour or 8.64000 E+4 for day

(exponent notation, here: 3.6 x10

3

or 8.64 x10

-4

).

Set factor FT in seconds.

Setting range • 1.00000 E-9 to 9.99999 E+9 (= 10

-9

to 10

+9

)

Press

↵

key to return to Fct. 3.05 “User Units”.


Advertising
This manual is related to the following products: