Part B IFC 210 Signal converter Sect. 5.18

05/2002

IFC 210 E

51

5.18 Application (Fct. 3.06)

Fct. 3.06 Application
Press

→

key.

Flow = Set characterization for the flow

•

steady

flow is steady, max. 150 % of Q

100%

•

pulsating

pulsating flow, = up to 1000 % of Q

100%

, e.g. due to

reciprocating pumps,

Select with

↑

or

↓

key.

see also Sect. 6.4 - 6.6.

Press

↵

key to return to Fct. 3.06 Application.

Empty pipe = signals that measuring tube is empty (possible only with built-in option)

•

No

•

Yes

Select with

↑

or

↓

key.

see also Sect. 6.4 - 6.6.
Press

↵

key to return to Fct. 3.06 Application.


5.19 Hardware (Fct. 3.07)

Assignment of terminals B1 + B2 and field current supply

Fct. 3.07 Hardware
Press

→

key.

Terminal B1 = define function of terminal B1 (valid for Fct. 1.07)

•

Statusoutput

•

Controlinput

Select with

↑

or

↓

key.

Press

↵

key to change to subfunction “Terminal B2”.

Terminal B2 = define function of terminal B2 (valid for Fct. 1.08)

•

Statusoutput

•

Controlinput

Select with

↑

or

↓

key.

Press

↵

key to change to subfunction “Fieldcurrent”.

Field current = define field current supply

•

Intern

•

Extern (> DN 1600 / > 64”)

Select with

↑

or

↓

key.

Press

↵

key to return to Fct. 3.07 Hardware.

Please note:
Only correct settings will give accurate measurement results.
The setting “external” is only required if a field current power driver is connected.