To Select Reference Temperature

This screen allows you to select the reference temperature to be
used in the determination of the conductivity of your sample.

1. Access the Select Reference Temperature screen from the

Conductivity Setup screen. The current Reference Temperature
is displayed.

2. Touch C or F to select the desired Temperature units for the

reference temperature.

3. Use the arrow keys to highlight the desired reference

temperature

4. Touch

enter to accept the reference temperature and return to

the conductivity Setup screen.

OR

Touch exit to return to the Conductivity Setup screen, without
making any changes.

The units of all temperature parameters must match. The meter will
automatically modify units of Default Temperature, Reference
Temperature and the Temperature Coefficient to match the last
choice made on any of these 3 screens. If Kelvin is desired, it must
be selected in either the default Temperature screen of the
Temperature Coefficient screen.