Instruction Manual

CyberScan PC5500/ 5000

55

This setup screen allows you to determine how quickly the meter will
respond to electrode drift. There are 3 speed settings: fast, medium
and slow.


To set pH Stability Criteria

1. Access the Set pH Stability Criteria screen from the pH (Ion)

Setup screen. The current stability criteria are displayed on the
screen.

2. Use

the

arrow keys to highlight the desired stability criteria.


3. Touch

enter to accept the stability criteria and return to the pH

(Ion) Setup screen and return to the pH Setup screen.

OR

Touch exit to return to the pH (Ion) Setup screen, without making
any changes.

Stability criteria are more stringent at the slower setting. Therefore, if
the highest precision is required, then a slow setting would be
desired. The default setting is the medium speed and this would be
adequate for the majority of applications.