Thermo Fisher Scientific CyberScan PC 5000 User Manual

Page 12

Advertising
background image

Instruction Manual

CyberScan PC5500/ 5000

7

3. Rinse and blot-dry (don’t wipe)
electrodes between each
measurement.

Rinse

electrodes

with distilled or deionized water,

or a portion of the next solution

to be measured.








4. Between

measurements,

store

conventional

pH

electrodes

storage solution, pH 4 buffer, or

KCI solution. Always leave the

filling hole of liquid filled

combination electrodes open.

Refill

when the level of

solution gets below the

manufacturer’s

recommended

level.


Proper electrode care is fundamental to obtaining reliable pH
measurements. Improper care of the electrode may cause the meter
reading to drift, respond slowly, or produce erroneous readings. For
this reason, the electrode should always be conditional and used in
accordance with manufacturer’s instructions.