To “standardize” your conductivity probe in the Advanced Level
of Operation
1. Access the conductivity Setup screen and verify the values for
the nominal cell constant of the probe, the reference temperature
and the temperature coefficient. Also, verify that the meter is set
for the uS/cm units.
2. Touch std on the Conductivity Measure screen to access the
Conductivity Standardization screen.
3. Immerse the conductivity and ATC probes in the standard with a
known conductivity value and stir.
4. Touch
clear to delete previous standardization values.
5. Touch
std to access the standardization mode again.
6. Use the keypad on the display screen to enter the value of the
known conductivity standard.
7. Touch
enter on the keypad to accept this conductivity standard
and return to the conductivity measure screen.
The actual cell constant will be displayed in the data box at the
bottom of the conductivity measure screen.
It is important to condition your conductivity probe according to the
manufacturer’s instructions prior to use.
You will only need one conductivity standard to standardize the
meter.