CyberScan PC5500/ 5000

64

6.12 Set Print Interval

Set pH Print Interval

Current INTERVAL

timed

enter

1 2 3

4 5 6

7 8 9

BS 0 .

manual


You have three options for setting the print interval: manual printing,
stable reading printing, and timed interval printing.

For manual printing of data

In this mode, data is printed only when you touch print on the pH
(mV, Ion, Conductivity) Measure screen.

1. Access the Set Print Interval screen from the pH (mV, Ion,

Conductivity) Setup screen. The current print interval is displayed
on the screen.

2. Touch

MAN to set the meter for manual printing.

3. Touch

enter to accept the print interval mode and return to the

pH (mV, Ion, Conductivity) Setup screen.

Printing is now done manually by touching print on the Measure
screen.

OR

Touch exit to return to the pH (mV, Ion, Conductivity) setup
screen, without making any changes.