3.2

Connection Diagram

Caution: Ensure electrical mains are disconnected before proceeding.

Connection for wall mounting version

RE LA RELB WASH RE LAY ALARM RELAY

-

+

Current
OP 2

29

28

19

GND +12V

-

+

Current
OP1

30

31

20

E

22

PT100/
PT1000

27

LIQ
GND

25

24

26

HOLD

14

13

12

9

10

11

8

7

N L

E

3

4

2

1

5

6

pH/ mV

1.

AC mains live wire

17.

no connection

2.

AC mains neutral wire

18.

no connection

3.

AC mains protective earth wire

19.

12V Power supply

4.

Alarm relay (NO)

20.

12V ground

5.

Alarm relay common

21.

no connection

6.

Alarm relay (NC)

22.

Earth ground

7.

Wash relay

23.

no connection

8.

Wash relay

24.

Temperature ground

9.

Relay B (SP 2)

25.

Temperature input

10.

Relay B (SP 2)

26.

Temperature sense (short to

11.

Relay A (SP 1)

terminal 25 if using 2- wire RTD)

12.

Relay A (SP 1)

27.

PMP, liquid ground in symmetrical

mode

13.

Hold function

28.

4-20 mA temperature output, -ve

terminal

14.

Hold function

29.

4-20 mA temperature output, +ve

terminal

15.

no connection

30.

4-20 mA pH output, -ve terminal

16.

no connection

31.

4-20 mA pH output, +ve terminal


IMPORTANT: The Alarm relay functions as an “Active Low” device i.e. it switches OFF under
Alarm condition. Therefore the Alarm display device should be connected to the ‘NC’ contacts
of the relay (5 & 6).