2 specifications for wall mount version, Pecifications for wall mount version – Thermo Fisher Scientific Alpha pH 2000 User Manual

Page 43

Advertising
background image

EMC Specifications

Emitted Interference

EN 61 326

Immunity to Interference

EN 61 326


Environmental Conditions
Ambient Temp. Operating
Range

-10 to 50

o

C (14 to 122

o

F)

Rel. Humidity

10 to 95% (non-condensing)

Power Supply

Input

80 to 250 VAC/DC 48/62 Hz

Main Fuse

250 mA anti-surge

Pollution Degree

2

Transient Overvoltage category

II

8.2

Specifications for wall mount version

Electrical Data and
Connections

Signal output

Two 0/4 to 20 mA outputs for pH/mv and
temperature , galvanically isolated

Load Max.600

pH / ORP Input

BNC (10

12

Impedance)

Connection terminal

(1 x 3-pins; 8-pin, 9-pin & 13-pin terminal blocks)

Mechanical Specification

Dimensions

144 x 144 x 111.5 mm (5.67 x 5.67 x 4.39 inch)

Weight

745g (unit); 1100g (boxed)

Material PBT
Insulation

NEMA 4X, IP 65