Thermo Fisher Scientific Alpha pH 2000 User Manual

Page 33

Advertising
background image

-

OFF = controller off

Use control Off to operate controller as a monitor only or to prevent relays

from switching.

-

L.CT = limit value control (on/off control).

Use limit control with pumps or values for fast response

-

P/PI = proportional/integral control

Use proportional control to operate your pumps smoothly or for precise
control of proportioning valves. Use PI controller to eliminate steady state
error.

Note: please refer to Appendix 4 for detailed information on controller
settings.

Press the ENT key to confirm your selection.

If the controller is set to limit value control (L.CT):

3. Selecting the relay status under Non-Alarm condition: press the

▲ or

▼ key to choose the desired relay status (de-energized = “DEEN” or
energized = “EN”). Press the ENT key to confirm your selection.

-

If the controller is set to proportional control (P/PI):

4. Selecting the proportional controller type: press the

▲ or ▼ key to

select the suitable controller type (“PLC” = pulse length control, “PFC” =
pulse frequency control).

Press the ENT key to confirm your selection.

If the proportional controller type is set to pulse length control (PLC):

5. Selecting the relay status under Non-Alarm condition: press the

▲ or

▼ key to choose the desired relay status (de-energized = “DEEN” or
energized = “EN”). Press the ENT key to confirm your selection.

6. Setting the proportional range. Press the

▲ or ▼ key to set the

proportional range (setting range: 10 to 500%). Press the ENT key to
confirm your setting.

7. Setting the pulse length. Press the

▲ or ▼ key to set the pulse length

(setting range: 0.5 to 20 seconds). Press the ENT key to confirm your
setting.

8. Setting the integral action time. Press the

▲ or ▼ key to set the

integral action time (setting range: 0.0 to 999.9 minutes). Press the ENT
key
to confirm your setting.

Note: If the integral action time is set to ‘0.0’, then the unit functions as a
‘P’ Controller.

If the proportional controller type is set to pulse frequency control (PFC):

9. Selecting the relay status under Non-Alarm condition: press the

▲ or

▼ key to choose the desired relay status (de-energized = “DEEN” or
energized = “EN”). Press the ENT key to confirm your selection.