Thermo Fisher Scientific Alpha pH 2000 User Manual
Page 31

the relay for the number of seconds (0 to 1999 seconds) you select. Press
the ENT key to confirm your setting.
Note: You can set a time delay for each relay, which stops the relay
from switching on the moment the set point is exceeded. This controller
lets you set a 0 to 1999 second time delay before your relay activates.
6. Setting the off-delay time lag: press the
▲ or ▼ key to enter the off-
delay time for set point 1 (set point 2). Your controller will delay
deactivation of the relay for the number of seconds (0 to 1999) you select.
Press the ENT key to confirm your setting.
Note: You can set a time delay for each relay, which stops the relay
from switching off the moment the value reaches the set point and
hysteresis. This controller lets you set a 0 to 1999 second time delay
before your relay deactivates.
Continue with Setup mode procedures, or return to Measurement mode by
pressing the
▲ and ▼ keys simultaneously (escape).