Thermo Fisher Scientific Alpha pH 2000 User Manual

Page 20

Advertising
background image

Connection for panel mounting version

1.

AC mains live wire

17.

no connection

2.

AC mains neutral wire

18.

no connection

3.

AC mains protective earth wire

19.

12V Power supply

4.

Alarm relay (NO)

20.

12V ground

5.

Alarm relay common

21.

no connection

6.

Alarm relay (NC)

22.

Earth ground

7.

Wash relay

23.

no connection

8.

Wash relay

24.

Temperature ground

9.

Relay B (SP 2)

25.

Temperature input

10.

Relay B (SP 2)

26.

Temperature sense (short to terminal 25 if

11.

Relay A (SP 1)

using 2- wire RTD

12.

Relay A (SP 1)

27.

PMP, liquid ground in symmetrical mode

13.

Hold function

28.

4-20 mA temperature output, -ve terminal

14.

Hold function

29

4-20 mA temperature output, +ve terminal

15.

no connection

30.

4-20 mA pH output, -ve terminal

16.

no connection

31.

4-20 mA pH output, +ve terminal

IMPORTANT: The Alarm relay functions as an “Active Low” device i.e. it switches OFF under
Alarm condition. Therefore the Alarm display device should be connected to the ‘NC’ contacts
of the relay (5 & 6).

NOTE:
Switch or circuit breaker shall included in the building installation.
It shall be in close proximity to the equipment and within easy reach of the
operator.
It shall be marked as the disconnecting device for the equipment.